Technical data
  • active diameter: 200 µm
  • sensitivity (typical): 1.0*10-7 V/Pa @5 MHz
  • max. pressure: 10 MPa
  • usable frequency range: 0.5-120 MHz
  • frequency response: +/-2 dB @1-40 MHz
  • membrane diameter: 80 mm
  • outer diameter: 130 mm
  • height: 15 mm
  • min. measure distance: 6 mm front, 9 mm back
Fields of application
  • standard sound field measurements
  • regulatory measurements according to IEC 62127 and 60601 for all medical Ultrasound devices
  • characterisation of NDT sensors
Users
  • certification laboratories
  • manufacturers of medical ultrasound devices
  • research and development laboratories
  • universities

Standard membrane hydrophone with backing MH-SB-0.2

VK-11026  
Technical data
  • active diameter: 200 µm
  • sensitivity (typical): 1.0*10-7 V/Pa @5 MHz
  • max. pressure: 10 MPa
  • usable frequency range: 0.5-120 MHz
  • frequency response: +/-2 dB @1-40 MHz
  • membrane diameter: 75 mm
  • outer diameter: 130 mm
  • height: 15 mm
  • min. measure distance: 2.5 mm, front only
Fields of application
  • continuous sound field measurements
  • field measurements with high speed positioning
  • quality measurements of ultrasound sensors
Users
  • certification laboratories
  • manufacturers of medical ultrasound devices

Standard membrane hydrophone MH-SD-0.4

VK-11041  
Technical data
  • active diameter: 400 µm
  • sensitivity (typical): 1.0*10-7 V/Pa @5 MHz
  • max. pressure: 10 MPa
  • usable frequency range: 0.5-40 MHz
  • frequency response: +/-2 dB @1-20 MHz
  • membrane diameter: 80 mm
  • outer diameter: 130 mm
  • height: 15 mm
  • min. measure distance: 6 mm front, 9 mm back
Fields of application
  • broadband measurements with low pressure amplitudes
  • regulatory measurements according to older standards for diagnostic ultrasound devices
User
  • certification laboratories
  • manufacturers of medical ultrasound devices

Standard membrane hydrophone MH-SD-0.5

   
   
   
   

Standard membrane hydrophone MH-SD-1.0

VK-11042  
Technical data
  • active diameter: 1000 µm
  • sensitivity (typical): 2.4*10-7 V/Pa @5 MHz
  • max. pressure: 10 MPa
  • usable frequency range: 0.5-30 MHz
  • frequency response: +/-4 dB @1-20 MHz
  • membrane diameter: 80 mm
  • outer diameter: 130 mm
  • height: 15 mm
  • min. measure distance: 6 mm front, 9 mm back
Fields of application
  • broadband measurements with very low pressure amplitudes
  • regulatory measurements according to older standards for diagnostic Ultrasound devices
  • measurements for acoustic material characterization (absorption measurement)
Users
  • certification laboratories
  • manufacturers of medical ultrasound devices
  • research labs
  • universities

Chemical capsulated membrane hydrophon MH-CC-0.4

VK-11031  
Technical data
  • active diameter: 325 µm
  • sensitivity (typical): 8.0*10-8 V/Pa @5 MHz
  • max. pressure: 20 MPa
  • usable frequency range: 0.5-30 MHz
  • frequency response: +/-4 dB @1-20 MHz
  • membrane diameter: 50 mm
  • outer diameter: 130 mm
  • height: 18 mm
  • min. measure distance: 8 mm front, 10 mm back
Fields fo application
  • measurements in aggressive liquids
  • development of sensors for chemical, petrol or food industries
  • development of flow sensors